WatchCare Smart Pad
Baxter International Inc. has announced it has issued an Urgent Medical Device Correction for the WatchCare Incontinence Management System due to potential for radio frequency (RF) interference with other medical devices. The WatchCare system is designed to discreetly alert the caregiver of an incontinence event. Although the WatchCare system has been developed to comply with the most recent RF standards, it radiates RF that might affect other devices in the vicinity, including devices on both patients and staff members. This RF interference could result in erroneous readings or additional malfunctions of these other devices and could therefore result in inappropriate medical intervention. Depending on the intended use of the device that malfunctions, there may be different hazardous situations that could occur. This product is manufactured by Hillrom, which was acquired by Baxter in late 2021.
Baxter is also issuing an updated customer notification letter for this correction, replacing the previous letter issued on September 30, 2022.
To date, interference in all cases but two is known to have occurred at distances less than one meter. There is insufficient data about distance on the remaining two reported interferences. Of note, Baxter is informing users of the following potential hazards, though additional devices may be affected:
Since this issue may affect blood glucose sensors, insulin pumps, fetal monitors, and general infusion pumps among other devices, serious harm or death may occur. This issue can affect medical devices on patients as well as staff caring for patients. No serious injuries or deaths have been reported.
Users and facilities should immediately locate all affected devices and stop use of all WatchCare system accessories where possible until this functionality is temporarily disabled and while Baxter continues to work to determine the cause of this problem. Baxter will contact users to arrange for WatchCare to be temporarily disabled.
Until all RF capabilities are disabled and/or all devices are removed from clinical care areas, be aware that RF emission from a functioning WatchCare device may potentially impact other devices (including, but not limited to, insulin pumps/blood glucose sensors, fetal monitor/dopplers, infusion pumps, telemetry devices, and bladder scanners). Please double-check all unexpected or atypical results and monitor infusions closely, if possible. Until this issue is resolved, use standard, non-RF-based, incontinence management pads.
Baxter is monitoring reports of RF interference and validating the impact the WatchCare device may have on other devices in the vicinity. Baxter is also investigating improvement opportunities and will provide a follow-up communication to users once available.
This Urgent Medical Device Correction applies to Centrella Bed with WatchCare product number P7900B; WatchCare System for VersaCare Bed Rev. A-J product number P00697901; WatchCare System for VersaCare Bed Rev. K product number P00697902; WatchCare System for Progressa Bed product number P00697903; and WatchCare System for Centrella Bed product number P00697905.
The WatchCare System is distributed in the United States.
Source: FDA
This link is being provided as a convenience and for informational purposes only it is not an endorsement or an approval of any of the products, services or opinions of the corporation or organization or individual.
Recallinsider.com bears no responsibility for the accuracy, legality or content of the external site or for that of subsequent links.
Customers with additional questions can contact Hillrom Technical Support by phone at 800-445-3720 or by email at technical.support@hillrom.com.
Any adverse events experienced with the use of this product may be reported using one of the following options:
This link is being provided as a convenience and for informational purposes only it is not an endorsement or an approval of any of the products, services or opinions of the corporation or organization or individual.
Recallinsider.com bears no responsibility for the accuracy, legality or content of the external site or for that of subsequent links.
This link is being provided as a convenience and for informational purposes only it is not an endorsement or an approval of any of the products, services or opinions of the corporation or organization or individual.
Recallinsider.com bears no responsibility for the accuracy, legality or content of the external site or for that of subsequent links.
This link is being provided as a convenience and for informational purposes only it is not an endorsement or an approval of any of the products, services or opinions of the corporation or organization or individual.
Recallinsider.com bears no responsibility for the accuracy, legality or content of the external site or for that of subsequent links.
Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA’s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
This link is being provided as a convenience and for informational purposes only it is not an endorsement or an approval of any of the products, services or opinions of the corporation or organization or individual.
Recallinsider.com bears no responsibility for the accuracy, legality or content of the external site or for that of subsequent links.
This link is being provided as a convenience and for informational purposes only it is not an endorsement or an approval of any of the products, services or opinions of the corporation or organization or individual.
Recallinsider.com bears no responsibility for the accuracy, legality or content of the external site or for that of subsequent links.
This link is being provided as a convenience and for informational purposes only it is not an endorsement or an approval of any of the products, services or opinions of the corporation or organization or individual.
Recallinsider.com bears no responsibility for the accuracy, legality or content of the external site or for that of subsequent links.
This link is being provided as a convenience and for informational purposes only it is not an endorsement or an approval of any of the products, services or opinions of the corporation or organization or individual.
Recallinsider.com bears no responsibility for the accuracy, legality or content of the external site or for that of subsequent links.
This link is being provided as a convenience and for informational purposes only it is not an endorsement or an approval of any of the products, services or opinions of the corporation or organization or individual.
Recallinsider.com bears no responsibility for the accuracy, legality or content of the external site or for that of subsequent links.